Analysis of the Systematic Errors Found in the Kipp & Zonen Large-Aperture Scintillometer

Studies have shown a systematic error in the Kipp & Zonen large-aperture scintillometer (K&ZLAS) measurements of the sensible heat flux, H. We improved on these studies and compared four K&ZLASs with a Wageningen large-aperture scintillometer at the Chilbolton Observatory. The scintillometers were installed such that their footprints were the same and independent flux measurements were made along the measurement path. This allowed us to compare H and the direct scintillometer output, the refractive index structure parameter, C2n. Furthermore, spectral analysis was performed on the raw scintillometer signal to investigate the characteristics of the error. Firstly, correlation coefficients =0.99 confirm the robustness of the scintillometer method, and secondly we discovered two systematic errors: the low-C2n error and the high-C2n error. The low-C2n error is a non-linear error that is caused by high-frequency noise, and we suspect the error to be caused by the calibration circuit in the receiver. It varies between each K&ZLAS, is significant for H = 50 W m-2, and we propose a solution to remove this error using the demodulated signal. The high-C2n error identified by us is the systematic error found in previous studies.We suspect this error to be caused by poor focal alignment of the receiver detector and the transmitter light-emitting diode that causes ineffective use of the Fresnel lens in the current Kipp & Zonen design. It varies between each K&ZLAS (35% up to 240%) and can only be removed by comparing with a reference scintillometer in the field.

Saved in:
Bibliographic Details
Main Authors: van Kesteren, A.J.H., Hartogensis, O.K.
Format: Article/Letter to editor biblioteca
Language:English
Subjects:area, c-n(2), heterogeneous land-surface, optical scintillometer, refractive-index, sensible heat fluxes,
Online Access:https://research.wur.nl/en/publications/analysis-of-the-systematic-errors-found-in-the-kipp-amp-zonen-lar
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Studies have shown a systematic error in the Kipp & Zonen large-aperture scintillometer (K&ZLAS) measurements of the sensible heat flux, H. We improved on these studies and compared four K&ZLASs with a Wageningen large-aperture scintillometer at the Chilbolton Observatory. The scintillometers were installed such that their footprints were the same and independent flux measurements were made along the measurement path. This allowed us to compare H and the direct scintillometer output, the refractive index structure parameter, C2n. Furthermore, spectral analysis was performed on the raw scintillometer signal to investigate the characteristics of the error. Firstly, correlation coefficients =0.99 confirm the robustness of the scintillometer method, and secondly we discovered two systematic errors: the low-C2n error and the high-C2n error. The low-C2n error is a non-linear error that is caused by high-frequency noise, and we suspect the error to be caused by the calibration circuit in the receiver. It varies between each K&ZLAS, is significant for H = 50 W m-2, and we propose a solution to remove this error using the demodulated signal. The high-C2n error identified by us is the systematic error found in previous studies.We suspect this error to be caused by poor focal alignment of the receiver detector and the transmitter light-emitting diode that causes ineffective use of the Fresnel lens in the current Kipp & Zonen design. It varies between each K&ZLAS (35% up to 240%) and can only be removed by comparing with a reference scintillometer in the field.