A compact miniaturized flow system based on low-temperature co-fired ceramic technology coupled to led mini-photometer for determination of dipyrone in pharmaceutical formulations.
In this work, an analytical microsystem based on LTCC (low-temperature co-fired ceramic) technology with monolithic incorporation of an optical flow cell for determination of dipyrone in pharmaceuticals is described. The detection system is based on the formation of a blue chromophore between dipyrone and Fe(III) photometrically monitored at 630 nm using a lab-made LED mini-photometer constructed with a light emitting diode as a radiation source and a Si photodiode as a detector. The lab-made mini-photometer elaborated presented a good performance in regard to high signal/noise ratio, low drift and good sensitivity. The analytical curve was linear in the dipyrone concentration range from 1.0 × 10-4 to 3.5 × 10-3 mol L-1, limit of detection of 4.5 × 10-5 mol L-1 and total time of analyses of 20 s yielding an analytical frequency of 195 h-1 with a low waste generation per analysis (480 µL).
Main Authors: | , , , , , , |
---|---|
Other Authors: | |
Format: | Artigo de periódico biblioteca |
Language: | Ingles English |
Published: |
2013-06-04
|
Subjects: | Low temperature, Co-fired ceramic, Analytical microsystems, Dipyrone, Flow injection, LED photometer, |
Online Access: | http://www.alice.cnptia.embrapa.br/alice/handle/doc/959237 https://doi.org/10.5935/0103-5053.20130112 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In this work, an analytical microsystem based on LTCC (low-temperature co-fired ceramic) technology with monolithic incorporation of an optical flow cell for determination of dipyrone in pharmaceuticals is described. The detection system is based on the formation of a blue chromophore between dipyrone and Fe(III) photometrically monitored at 630 nm using a lab-made LED mini-photometer constructed with a light emitting diode as a radiation source and a Si photodiode as a detector. The lab-made mini-photometer elaborated presented a good performance in regard to high signal/noise ratio, low drift and good sensitivity. The analytical curve was linear in the dipyrone concentration range from 1.0 × 10-4 to 3.5 × 10-3 mol L-1, limit of detection of 4.5 × 10-5 mol L-1 and total time of analyses of 20 s yielding an analytical frequency of 195 h-1 with a low waste generation per analysis (480 µL). |
---|