Evaluation of the accuracy of various phenotypic tests to detect oxacillin resistance in coagulase-negative staphylococci

In this study, we determined the accuracy of phenotypic tests (cefoxitin agar dilution, 30µg-cefoxitin and 1µg-oxacillin disks) to detect the oxacillin resistance/mecA gene among coagulase-negative staphylococci (CoNS) isolates. The presence of the mecA gene was detected by PCR technique (gold standard). A total of 176 CoNS isolates from blood of hospitalized patients were evaluated. Of these, 138 (78.4%) harbored the mecA gene. Using 30µg-cefoxitin and 1µg-oxacillin disks we obtained 100 and 98.3% accuracy, respectively. In addition, when cefoxitin was used as marker in an agar dilution method, the higher accuracy (99.4%) was established with 8mg cefoxitin per mL breakpoint. Thus, despite of the agar dilution method using cefoxitin as a marker not being standard for this detection, our results suggested that it is an excellent alternative to detect the oxacillin resistance/mecA gene among CoNS isolates.

Saved in:
Bibliographic Details
Main Authors: Perez,Leandro Reus Rodrigues, d'Azevedo,Pedro Alves
Format: Digital revista
Language:English
Published: Brazilian Society of Infectious Diseases 2008
Online Access:http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1413-86702008000300009
Tags: Add Tag
No Tags, Be the first to tag this record!