Biomechanical evaluation of a new femoral stem design for total hip replacement in a canine model
ABSTRACT Purpose To evaluate the biomechanical properties of a novel total hip replacement femoral stem. Methods Eight pairs of femurs from dog cadavers were used. The femurs were separated into different groups. A novel femoral stem with a convex proximal portion (Stem B) was biomechanically evaluated and compared to awell-known veterinary collared stem (Stem A). Femoral stems were inserted into the contralateral femurs from the same dog, forming 16 constructs. A flexo-compression load was applied on the axial axis of each sample. Maximum strength, deflection, stiffness, and energy absorption were analysed. Results Group B constructs showed significantly higher values (p ? 0.05) for the variables, except stiffness. The mean maximum strength was 1,347 ± 357 N for Group A and 1,805 ± 123 N for Group B (p ? 0.0069). The mean deflection was5.54 ± 2.63 mm for Group A and 10.03 ± 3.99 mm for Group B (p ? 0.0056). For the energy variable, the force was 6,203 ± 3,488 N/mm for Group A and 12,885 ± 5,056 N/mm for Group B (p ? 0.0054). Stem B had greater maximum strength, deflection, and energy. Conclusions The new stem was effective in neutralizing the impact of axial flexion-compression stresses during biomechanical tests in cadaveric models.
Main Authors: | , , , , , , , , , , |
---|---|
Format: | Digital revista |
Language: | English |
Published: |
Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia
2021
|
Online Access: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-86502021000500204 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|