PROCEDIMENTO AUTOMÁTICO PARA DETERMINAÇÃO ESPECTROFOTOMÉTRICA DE HIDROQUINONA EMPREGANDO MULTICOMUTAÇÃO EM SISTEMA DE ANÁLISE EM FLUXO

This paper describes a procedure based on multicommutation in flow analysis system for photometric determination of hydroquinone in pharmaceutical gel formulations. The procedure involves oxidation of hydroquinone by Fe (III) and subsequent reaction between the generated Fe (II) species with 1,10-phenanthroline, producing a complex detected at 510 nm. Under optimized conditions, the proposed system showed a linear response in the range between 5 to 120 µmol L-1 (r = 0.9995), a limit of detection (3σ) 0.52 µmol L-1 and a relative standard deviation (RSD) of 1.5% for a 80 µmol L-1 hydroquinone standard solution. Others profitable features such as an analytical frequency of 72 samples per hour; a reagent consumption of 0.432 and 0.117 mg 1-10-phenantroline and FeCl3.6H2O per determination, respectively; and a waste generation of 4.0 ml per determination were also achieved.

Saved in:
Bibliographic Details
Main Authors: Lima,Manoel de Jesus de Aquino, Castilho,Quésia Guedes da Silva, Fernandes,Ridvan Nunes, Tanaka,Auro Atsushi, Reis,Boaventura Freire dos
Format: Digital revista
Language:Portuguese
Published: Sociedade Brasileira de Química 2016
Online Access:http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422016000500592
Tags: Add Tag