Simultaneous and accurate determination of water- and fat-soluble vitamins in multivitamin tablets by using an RP-HPLC method

In the present study, a reversed-phase high-performance liquid chromatographic (RP-HPLC) procedure was developed and validated for the simultaneous determination of seven water-soluble vitamins (thiamine, riboflavin, niacin, cyanocobalamin, ascorbic acid, folic acid, and p-aminobenzoic acid) and four fat-soluble vitamins (retinol acetate, cholecalciferol, α-tocopherol, and phytonadione) in multivitamin tablets. The linearity of the method was excellent (R² > 0.999) over the concentration range of 10 - 500 ng mL-1. The statistical evaluation of the method was carried out by performing the intra- and inter-day precision. The accuracy of the method was tested by measuring the average recovery; values ranged between 87.4% and 98.5% and were acceptable quantitative results that corresponded with the label claims.

Saved in:
Bibliographic Details
Main Authors: Kucukkolbasi,Semahat, Bilber,Onur, Ayyildiz,H. Filiz, Kara,Huseyin
Format: Digital revista
Language:English
Published: Sociedade Brasileira de Química 2013
Online Access:http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422013000700020
Tags: Add Tag
No Tags, Be the first to tag this record!