Quantificação sérica de vitamina C por CLAE-UV e estudo de estabilidade

Vitamin C, an exogenous antioxidant, is essential to human health. In this study, a method was validated to serum vitamin C quantification by HPLC-UV. Its stability with and without the use of tris [2-carboxy-ethyl] phosphine hydrochloride (TCEP), at -20 and -80 °C, in serum and supernatant were also evaluated. Analysis showed r² &gt; 0.99, precision CV% < 15% and % bias < 15%, being linear, precise and accurate. The stability test revealed that using TCEP in serum storage at -20 and -80 °C or in supernatant at -80 °C the vitamin C levels remain stable for 30 and 12 days, respectively.