Response matrix of a multisphere neutron spectrometer with an ³He proportional counter

The response matrix of a Bonner sphere spectrometer was calculated by use of the MCNP code. As thermal neutron counter, the spectrometer has a 3.2 cm-diameter ³He-filled proportional counter which is located at the center of a set of polyethylene spheres. The response was calculated for 0, 3, 5, 6, 8, 10, 12, and 16 inches-diameter polyethylene spheres for neutrons whose energy goes from 10~9 to 20 MeV. The response matrix was compared with a set of responses measured with several monoenergetic neutron sources. In this comparison the calculated matrix agrees with the experimental results. The matrix was also compared with the response matrix calculated for the PTB C spectrometer. Even though that calculation was carried out using a detailed model to describe the proportional counter; both matrices do agree, but small differences are observed in the bare case because of the difference in the model used during calculations. Other differences are in some spheres for 14.8 and 20 MeV neutrons, probably due to the differences in the cross sections used during both calculations.

Saved in:
Bibliographic Details
Main Authors: Vega-Carrillo,H.R., Manzanares-Acuña,E., Hernández-Dávila,V Martín, Mercado Sánchez,G.A.
Format: Digital revista
Language:English
Published: Sociedad Mexicana de Física 2005
Online Access:http://www.scielo.org.mx/scielo.php?script=sci_arttext&pid=S0035-001X2005000100008
Tags: Add Tag
No Tags, Be the first to tag this record!