Determinação simultânea de cátions empregando eletroforese capilar com detecção condutométrica sem contato em equipamento construído em laboratório
This work describes the development of a home-made capillary electrophoresis (CE) system based on the capacitively coupled contactless conductivity detection (C4D) for the separation of the metallic species Zn2+, Cr3+, Pb2+, Cd2+, Co2+, Cu2+, Ni2+ e Tl+. A background electrolyte composed of MES/Histidine 0,02 mol L-1 (pH 5.0) was optimized for the separation of the metallic species by using organic solvents and complexing agents as additives. The system allowed the determination of the metallic species using MES/Histidine 0,02 mol L-1 and methanol 5% (pH 5.0) as a background electrolyte, 15 kV separation voltage and hydrodynamic injection by gravity.
Main Authors: | , , , , , , |
---|---|
Format: | Digital revista |
Language: | Portuguese |
Published: |
Sociedade Brasileira de Química
2009
|
Online Access: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422009000800036 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | This work describes the development of a home-made capillary electrophoresis (CE) system based on the capacitively coupled contactless conductivity detection (C4D) for the separation of the metallic species Zn2+, Cr3+, Pb2+, Cd2+, Co2+, Cu2+, Ni2+ e Tl+. A background electrolyte composed of MES/Histidine 0,02 mol L-1 (pH 5.0) was optimized for the separation of the metallic species by using organic solvents and complexing agents as additives. The system allowed the determination of the metallic species using MES/Histidine 0,02 mol L-1 and methanol 5% (pH 5.0) as a background electrolyte, 15 kV separation voltage and hydrodynamic injection by gravity. |
---|