Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets
ABSTRACT Objective: To investigate infusion pumps start-up delay according to different brands of infusion pumps, flow rates and intravenous sets priming techniques. Method: The experimental study simulated clinical practice under controlled conditions, using a 50 mL syringe with NaCl 0.9% solution, two syringe infusion pumps (A and B), six rates (0.3, 0.5, 1.0, 5, 10 and 20 mL/h), two purging techniques (manually or infusion pump’s electronic bolus). Data were analyzed according to mean, standard deviation, Student’s t and ANOVA tests (p<0.05). Results: The start-up delay was greater in low rates regardless the priming technique. The electronic bolus increased the infusion pump A accuracy at 0.3mL/h (p=0.010), 0.5 mL/h (p=0.002) and 1.0mL/h (p=0.004). Pump’s accuracy in all studied rates and manual IV sets filling was similar. Conclusion: In low infusion rates the start-up delay was greater despite the infusion pump brand and electronic bolus improved pumps accuracy.
Main Authors: | , , , , , |
---|---|
Format: | Digital revista |
Language: | English |
Published: |
Universidade Federal do Rio Grande do Sul. Escola de Enfermagem
2022
|
Online Access: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1983-14472022000100419 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
oai:scielo:S1983-14472022000100419 |
---|---|
record_format |
ojs |
spelling |
oai:scielo:S1983-144720220001004192022-05-19Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust setsVieira,Natalie MarinoPires,Maria Paula de OliveiraCrespo,Gabriela BeltranNascimento,Larissa Perez PardoPeterlini,Maria Angélica SorginiPedreira,Mavilde Luz Gonçalves Critical care Infusion pumps Nursing Patient safety Syringes ABSTRACT Objective: To investigate infusion pumps start-up delay according to different brands of infusion pumps, flow rates and intravenous sets priming techniques. Method: The experimental study simulated clinical practice under controlled conditions, using a 50 mL syringe with NaCl 0.9% solution, two syringe infusion pumps (A and B), six rates (0.3, 0.5, 1.0, 5, 10 and 20 mL/h), two purging techniques (manually or infusion pump’s electronic bolus). Data were analyzed according to mean, standard deviation, Student’s t and ANOVA tests (p<0.05). Results: The start-up delay was greater in low rates regardless the priming technique. The electronic bolus increased the infusion pump A accuracy at 0.3mL/h (p=0.010), 0.5 mL/h (p=0.002) and 1.0mL/h (p=0.004). Pump’s accuracy in all studied rates and manual IV sets filling was similar. Conclusion: In low infusion rates the start-up delay was greater despite the infusion pump brand and electronic bolus improved pumps accuracy.info:eu-repo/semantics/openAccessUniversidade Federal do Rio Grande do Sul. Escola de EnfermagemRevista Gaúcha de Enfermagem v.43 20222022-01-01info:eu-repo/semantics/articletext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1983-14472022000100419en10.1590/1983-1447.2022.20210071.en |
institution |
SCIELO |
collection |
OJS |
country |
Brasil |
countrycode |
BR |
component |
Revista |
access |
En linea |
databasecode |
rev-scielo-br |
tag |
revista |
region |
America del Sur |
libraryname |
SciELO |
language |
English |
format |
Digital |
author |
Vieira,Natalie Marino Pires,Maria Paula de Oliveira Crespo,Gabriela Beltran Nascimento,Larissa Perez Pardo Peterlini,Maria Angélica Sorgini Pedreira,Mavilde Luz Gonçalves |
spellingShingle |
Vieira,Natalie Marino Pires,Maria Paula de Oliveira Crespo,Gabriela Beltran Nascimento,Larissa Perez Pardo Peterlini,Maria Angélica Sorgini Pedreira,Mavilde Luz Gonçalves Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets |
author_facet |
Vieira,Natalie Marino Pires,Maria Paula de Oliveira Crespo,Gabriela Beltran Nascimento,Larissa Perez Pardo Peterlini,Maria Angélica Sorgini Pedreira,Mavilde Luz Gonçalves |
author_sort |
Vieira,Natalie Marino |
title |
Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets |
title_short |
Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets |
title_full |
Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets |
title_fullStr |
Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets |
title_full_unstemmed |
Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets |
title_sort |
start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets |
description |
ABSTRACT Objective: To investigate infusion pumps start-up delay according to different brands of infusion pumps, flow rates and intravenous sets priming techniques. Method: The experimental study simulated clinical practice under controlled conditions, using a 50 mL syringe with NaCl 0.9% solution, two syringe infusion pumps (A and B), six rates (0.3, 0.5, 1.0, 5, 10 and 20 mL/h), two purging techniques (manually or infusion pump’s electronic bolus). Data were analyzed according to mean, standard deviation, Student’s t and ANOVA tests (p<0.05). Results: The start-up delay was greater in low rates regardless the priming technique. The electronic bolus increased the infusion pump A accuracy at 0.3mL/h (p=0.010), 0.5 mL/h (p=0.002) and 1.0mL/h (p=0.004). Pump’s accuracy in all studied rates and manual IV sets filling was similar. Conclusion: In low infusion rates the start-up delay was greater despite the infusion pump brand and electronic bolus improved pumps accuracy. |
publisher |
Universidade Federal do Rio Grande do Sul. Escola de Enfermagem |
publishDate |
2022 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1983-14472022000100419 |
work_keys_str_mv |
AT vieiranataliemarino startupdelayinsyringeinfusionpumpswithdifferentratesandprimingtechniquesofintravenoustsets AT piresmariapauladeoliveira startupdelayinsyringeinfusionpumpswithdifferentratesandprimingtechniquesofintravenoustsets AT crespogabrielabeltran startupdelayinsyringeinfusionpumpswithdifferentratesandprimingtechniquesofintravenoustsets AT nascimentolarissaperezpardo startupdelayinsyringeinfusionpumpswithdifferentratesandprimingtechniquesofintravenoustsets AT peterlinimariaangelicasorgini startupdelayinsyringeinfusionpumpswithdifferentratesandprimingtechniquesofintravenoustsets AT pedreiramavildeluzgoncalves startupdelayinsyringeinfusionpumpswithdifferentratesandprimingtechniquesofintravenoustsets |
_version_ |
1756436008579104768 |