Percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasound

ABSTRACT Purpose: To evaluate the efficiency of a novel device coupled with ultrassound for renal percutaneous puncture. Materials and Methods: After establishing hydronephrosis, ten pigs had three calyxes of each kidney punctured by the same urology resident, with and without the new device ("Punctiometer"). Time for procedure completion, number of attempts to reach the calyx, puncture precision and puncture complications were recorded in both groups and compared. Results: Puncture success on the first attempt was achieved in 25 punctures (83%) with the Punctiometer and in 13 punctures (43%) without the Punctiometer (p=0.011). The mean time required to perform three punctures in each kidney was 14.5 minutes with the Punctiometer and 22.4 minutes without the Punctiometer (p=0.025). The only complications noted were renal hematomas. In the Punctiometer group, all kidneys had small hematomas. In the no Punctiometer group 80% had small hematomas, 10% had a medium hematoma and 10% had a big hematoma. There was no difference in complications between both groups. Conclusions: The Punctiometer is an effective device to increase the likelihood of an accurate renal calyx puncture during PCNL, with a shorter time required to perform the procedure.

Saved in:
Bibliographic Details
Main Authors: Chan,Chen Jen, Srougi,Victor, Tanno,Fabio Yoshiaki, Jordão,Ricardo Duarte, Srougi,Miguel
Format: Digital revista
Language:English
Published: Sociedade Brasileira de Urologia 2015
Online Access:http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1677-55382015000500953
Tags: Add Tag
No Tags, Be the first to tag this record!
id oai:scielo:S1677-55382015000500953
record_format ojs
spelling oai:scielo:S1677-553820150005009532015-11-27Percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasoundChan,Chen JenSrougi,VictorTanno,Fabio YoshiakiJordão,Ricardo DuarteSrougi,Miguel Punctiometer percutaneous access kidney punction nephrolitotripsy ultrassound ABSTRACT Purpose: To evaluate the efficiency of a novel device coupled with ultrassound for renal percutaneous puncture. Materials and Methods: After establishing hydronephrosis, ten pigs had three calyxes of each kidney punctured by the same urology resident, with and without the new device ("Punctiometer"). Time for procedure completion, number of attempts to reach the calyx, puncture precision and puncture complications were recorded in both groups and compared. Results: Puncture success on the first attempt was achieved in 25 punctures (83%) with the Punctiometer and in 13 punctures (43%) without the Punctiometer (p=0.011). The mean time required to perform three punctures in each kidney was 14.5 minutes with the Punctiometer and 22.4 minutes without the Punctiometer (p=0.025). The only complications noted were renal hematomas. In the Punctiometer group, all kidneys had small hematomas. In the no Punctiometer group 80% had small hematomas, 10% had a medium hematoma and 10% had a big hematoma. There was no difference in complications between both groups. Conclusions: The Punctiometer is an effective device to increase the likelihood of an accurate renal calyx puncture during PCNL, with a shorter time required to perform the procedure.info:eu-repo/semantics/openAccessSociedade Brasileira de UrologiaInternational braz j urol v.41 n.5 20152015-10-01info:eu-repo/semantics/articletext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1677-55382015000500953en10.1590/S1677-5538.IBJU.2014.0586
institution SCIELO
collection OJS
country Brasil
countrycode BR
component Revista
access En linea
databasecode rev-scielo-br
tag revista
region America del Sur
libraryname SciELO
language English
format Digital
author Chan,Chen Jen
Srougi,Victor
Tanno,Fabio Yoshiaki
Jordão,Ricardo Duarte
Srougi,Miguel
spellingShingle Chan,Chen Jen
Srougi,Victor
Tanno,Fabio Yoshiaki
Jordão,Ricardo Duarte
Srougi,Miguel
Percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasound
author_facet Chan,Chen Jen
Srougi,Victor
Tanno,Fabio Yoshiaki
Jordão,Ricardo Duarte
Srougi,Miguel
author_sort Chan,Chen Jen
title Percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasound
title_short Percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasound
title_full Percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasound
title_fullStr Percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasound
title_full_unstemmed Percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasound
title_sort percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasound
description ABSTRACT Purpose: To evaluate the efficiency of a novel device coupled with ultrassound for renal percutaneous puncture. Materials and Methods: After establishing hydronephrosis, ten pigs had three calyxes of each kidney punctured by the same urology resident, with and without the new device ("Punctiometer"). Time for procedure completion, number of attempts to reach the calyx, puncture precision and puncture complications were recorded in both groups and compared. Results: Puncture success on the first attempt was achieved in 25 punctures (83%) with the Punctiometer and in 13 punctures (43%) without the Punctiometer (p=0.011). The mean time required to perform three punctures in each kidney was 14.5 minutes with the Punctiometer and 22.4 minutes without the Punctiometer (p=0.025). The only complications noted were renal hematomas. In the Punctiometer group, all kidneys had small hematomas. In the no Punctiometer group 80% had small hematomas, 10% had a medium hematoma and 10% had a big hematoma. There was no difference in complications between both groups. Conclusions: The Punctiometer is an effective device to increase the likelihood of an accurate renal calyx puncture during PCNL, with a shorter time required to perform the procedure.
publisher Sociedade Brasileira de Urologia
publishDate 2015
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1677-55382015000500953
work_keys_str_mv AT chanchenjen percutaneouspunctureofrenalcalyxesguidedbyanoveldevicecoupledwithultrasound
AT srougivictor percutaneouspunctureofrenalcalyxesguidedbyanoveldevicecoupledwithultrasound
AT tannofabioyoshiaki percutaneouspunctureofrenalcalyxesguidedbyanoveldevicecoupledwithultrasound
AT jordaoricardoduarte percutaneouspunctureofrenalcalyxesguidedbyanoveldevicecoupledwithultrasound
AT srougimiguel percutaneouspunctureofrenalcalyxesguidedbyanoveldevicecoupledwithultrasound
_version_ 1756428182847750144