Evaluation of Polyethylene Wear in a Brazilian Ultracongruent Knee Prosthesis with a Rotating Platform
Abstract Objective To evaluate the wear of polyethylene in a Brazilian ultracongruent knee prosthesis with a rotating platform (Rotaflex, Víncula, Rio Claro, SP, Brasil). Methods We used the test method with the loading and preparation parameters mentioned in the standards regulation ISO 14243-1:2009, and the measurement methods mentioned in the standards regulation ISO 14243-2:2009, for the evaluation of the wear behavior of a Brazilian prosthesis with a rotating platform. The equipment used for the wear test was the ISO 14243–1 gait simulator (EndoLab, Riedering, Germany). Results After 10 million cycles, the evaluation of the polyethylene wear showed a regular appearance of surface wear at a mean rate of 2.56 mg per million cycles. Conclusion The wear of the polyethylene of the evaluated prosthesis was minimal after the tests performed and with safety limits higher than those recommended by biomechanical engineering.
Main Authors: | Pécora,José Ricardo, Romero,Valéria |
---|---|
Format: | Digital revista |
Language: | English |
Published: |
Sociedade Brasileira de Ortopedia e Traumatologia
2021
|
Online Access: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-36162021000100042 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Similar Items
-
THE RESULTS OF A COMPARISON OF A BRAZILIAN AND AN IMPORTED KNEE PROSTHESIS - 5 YEARS OF FOLLOW-UP
by: CALEGARI,JOAO HENRIQUE COSTA, et al.
Published: (2022) -
REVISION TOTAL KNEE ARTHROPLASTY USING THE MODERN CONSTRAINED CONDYLAR KNEE PROSTHESIS
by: Nakano,Naoki, et al.
Published: (2016) -
Effect of singing training on total laryngectomees wearing a tracheoesophageal voice prosthesis
by: Onofre,Fernanda, et al.
Published: (2013) -
On the mechanical design of cruciate ligaments based on polymeric ropes for knee prosthesis
by: Rodriguez Lelis,JM, et al.
Published: (2011) -
Restless legs syndrome in subjects with a knee prosthesis: evidence that symptoms are generated in the periphery
by: Pereira Júnior,José Carlos, et al.
Published: (2011)