Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera
Purpose: To compare the results of central corneal thickness (CCT) measurements obtained using optical low-coherence reflectometry (OLCR), Fourier domain optical coherence tomography (FD-OCT), and a Scheimpflug camera (SC), combined with Placido corneal topography. Methods: A total of 25 healthy subjects were enrolled in the present study, and one eye of each subject was included. A detailed ophthalmic examination was performed in all cases following CCT measurements with OLCR, FD-OCT, and SC. The results were compared using an ANOVA test. Bland-Altman analysis was used to demonstrate agreement between methods. Intra-examiner repeatability was assessed by using intraclass correlation coefficients (ICCs). Results: Statistically significant differences were observed between the results of the CCT measurements obtained using the three different devices (p=0.009). Significant correlations were found between OLCR and FD-OCT (r=0.97; p<0.0001), FD-OCT and SC (r=0.91; p<0.0001), and OLCR and SC (r=0.95; p<0.0001). The 95% limits of agreement (LOA) obtained from Bland-Altman plots were from -7.2 μm to 28.7 μm for OLCR versus FD-OCT, from -19.2 μm to 30.4 μm for OLCR versus SC, and from -42.6 to 32.3 μm for FD-OCT versus SC. Intra-examiner repeatability was excellent for each method, with ICCs >0.98. Conclusion: Although the results of CCT measurements obtained from these three devices were highly correlated with one another and the mean differences between instruments were comparable with the reported diurnal CCT fluctuation, the measurements are not directly interchangeable in clinical practice because of the wide LOA values.
Main Authors: | , , , |
---|---|
Format: | Digital revista |
Language: | English |
Published: |
Conselho Brasileiro de Oftalmologia
2014
|
Online Access: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492014000600006 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
oai:scielo:S0004-27492014000600006 |
---|---|
record_format |
ojs |
spelling |
oai:scielo:S0004-274920140006000062015-04-10Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug cameraGonul,SabanKoktekir,Bengu EkinciBakbak,BerkerGedik,Sansal Cornea/anatomy & histology Corneal topography/methods Tomography, optical coherence/methods Fourier analysis Comparative study Observer variation Purpose: To compare the results of central corneal thickness (CCT) measurements obtained using optical low-coherence reflectometry (OLCR), Fourier domain optical coherence tomography (FD-OCT), and a Scheimpflug camera (SC), combined with Placido corneal topography. Methods: A total of 25 healthy subjects were enrolled in the present study, and one eye of each subject was included. A detailed ophthalmic examination was performed in all cases following CCT measurements with OLCR, FD-OCT, and SC. The results were compared using an ANOVA test. Bland-Altman analysis was used to demonstrate agreement between methods. Intra-examiner repeatability was assessed by using intraclass correlation coefficients (ICCs). Results: Statistically significant differences were observed between the results of the CCT measurements obtained using the three different devices (p=0.009). Significant correlations were found between OLCR and FD-OCT (r=0.97; p<0.0001), FD-OCT and SC (r=0.91; p<0.0001), and OLCR and SC (r=0.95; p<0.0001). The 95% limits of agreement (LOA) obtained from Bland-Altman plots were from -7.2 μm to 28.7 μm for OLCR versus FD-OCT, from -19.2 μm to 30.4 μm for OLCR versus SC, and from -42.6 to 32.3 μm for FD-OCT versus SC. Intra-examiner repeatability was excellent for each method, with ICCs >0.98. Conclusion: Although the results of CCT measurements obtained from these three devices were highly correlated with one another and the mean differences between instruments were comparable with the reported diurnal CCT fluctuation, the measurements are not directly interchangeable in clinical practice because of the wide LOA values. info:eu-repo/semantics/openAccessConselho Brasileiro de OftalmologiaArquivos Brasileiros de Oftalmologia v.77 n.6 20142014-12-01info:eu-repo/semantics/articletext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492014000600006en10.5935/0004-2749.20140087 |
institution |
SCIELO |
collection |
OJS |
country |
Brasil |
countrycode |
BR |
component |
Revista |
access |
En linea |
databasecode |
rev-scielo-br |
tag |
revista |
region |
America del Sur |
libraryname |
SciELO |
language |
English |
format |
Digital |
author |
Gonul,Saban Koktekir,Bengu Ekinci Bakbak,Berker Gedik,Sansal |
spellingShingle |
Gonul,Saban Koktekir,Bengu Ekinci Bakbak,Berker Gedik,Sansal Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera |
author_facet |
Gonul,Saban Koktekir,Bengu Ekinci Bakbak,Berker Gedik,Sansal |
author_sort |
Gonul,Saban |
title |
Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera |
title_short |
Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera |
title_full |
Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera |
title_fullStr |
Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera |
title_full_unstemmed |
Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera |
title_sort |
comparison of central corneal thickness measurements using optical low-coherence reflectometry, fourier domain optical coherence tomography, and scheimpflug camera |
description |
Purpose: To compare the results of central corneal thickness (CCT) measurements obtained using optical low-coherence reflectometry (OLCR), Fourier domain optical coherence tomography (FD-OCT), and a Scheimpflug camera (SC), combined with Placido corneal topography. Methods: A total of 25 healthy subjects were enrolled in the present study, and one eye of each subject was included. A detailed ophthalmic examination was performed in all cases following CCT measurements with OLCR, FD-OCT, and SC. The results were compared using an ANOVA test. Bland-Altman analysis was used to demonstrate agreement between methods. Intra-examiner repeatability was assessed by using intraclass correlation coefficients (ICCs). Results: Statistically significant differences were observed between the results of the CCT measurements obtained using the three different devices (p=0.009). Significant correlations were found between OLCR and FD-OCT (r=0.97; p<0.0001), FD-OCT and SC (r=0.91; p<0.0001), and OLCR and SC (r=0.95; p<0.0001). The 95% limits of agreement (LOA) obtained from Bland-Altman plots were from -7.2 μm to 28.7 μm for OLCR versus FD-OCT, from -19.2 μm to 30.4 μm for OLCR versus SC, and from -42.6 to 32.3 μm for FD-OCT versus SC. Intra-examiner repeatability was excellent for each method, with ICCs >0.98. Conclusion: Although the results of CCT measurements obtained from these three devices were highly correlated with one another and the mean differences between instruments were comparable with the reported diurnal CCT fluctuation, the measurements are not directly interchangeable in clinical practice because of the wide LOA values. |
publisher |
Conselho Brasileiro de Oftalmologia |
publishDate |
2014 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492014000600006 |
work_keys_str_mv |
AT gonulsaban comparisonofcentralcornealthicknessmeasurementsusingopticallowcoherencereflectometryfourierdomainopticalcoherencetomographyandscheimpflugcamera AT koktekirbenguekinci comparisonofcentralcornealthicknessmeasurementsusingopticallowcoherencereflectometryfourierdomainopticalcoherencetomographyandscheimpflugcamera AT bakbakberker comparisonofcentralcornealthicknessmeasurementsusingopticallowcoherencereflectometryfourierdomainopticalcoherencetomographyandscheimpflugcamera AT gediksansal comparisonofcentralcornealthicknessmeasurementsusingopticallowcoherencereflectometryfourierdomainopticalcoherencetomographyandscheimpflugcamera |
_version_ |
1756373195007459328 |