Comparative analysis of the nuclear lens opalescence by the Lens Opacities Classification System III with nuclear density values provided by Oculus Pentacam: a cross-section study using Pentacam Nucleus Staging software
PURPOSE: To compare the clinical classification of cataract using the Lens Opacities Classification System (LOCS) III with the mean values of lens density provided by the Pentacam Scheimpflug System in nuclear cataracts. METHODS: One hundred and one eyes from 101 patients with age-related nuclear cataract were submitted to clinical examination for lens grading score using LOCS III. According to LOCS III, nuclear opalescence was divided in six groups. Patients were evaluated by the Pentacam Scheimpflug System for the mean lens density using the Pentacam lens densitometry program (PLDP), the Pentacam Nucleus Staging (PNS) mean value and the PNS cataract grading score. RESULTS: A positive correlation between the mean values of lens density and LOCS III classification, considering groups 1 to 5, could be noticed with PLDP and PNS mean value. The mean values between the groups were similar using the PLDP and the PNS mean value. However, when the PNS cataract grading score was evaluated, there was low correspondence with LOCS III classification. CONCLUSION: Pentacam Scheimpflug device offers an objective measure of the lens nuclear density on nuclear cataracts. PLDP and the PNS mean value were both useful to evaluate age-related nuclear cataract up to LOCS III group 5.
Main Authors: | Magalhães,Fernanda Pedreira, Costa,Elaine Fiod, Cariello,Angelino Júlio, Rodrigues,Eduardo Buchele, Hofling-Lima,Ana Luisa |
---|---|
Format: | Digital revista |
Language: | English |
Published: |
Conselho Brasileiro de Oftalmologia
2011
|
Online Access: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492011000200008 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Similar Items
-
Comparing the postoperative refractive predictability of Pentacam HR and IOLMaster 500 after a multifocal intraocular lens implantation
by: Andrade Junior,Newton, et al.
Published: (2020) -
Repeatability of central corneal thickness measurement with the Pentacam HR system
by: Alonso,Ruiz Simonato, et al.
Published: (2012) -
PRK em olhos com topografia atípica e pentacam normal
by: Alves,Ermano de Melo, et al.
Published: (2013) -
Case-control study for clinical findings on intraocular lens opacity
by: Jorge,Priscilla de Almeida, et al.
Published: (2014) -
Effects of 1% cyclopentolate hydrochloride on anterior segment parameters obtained with Pentacam in young adults
by: Arici,Ceyhun, et al.
Published: (2014)