Evaluation of inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations.

The analytical performance of two inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations, equipped with charge coupled device solid-state detectors was evaluated using Ar, Ba, Mg, and Ni as test elements. Both instruments have similar Echelle optical arrangements and radio-frequency generators, differing only in the viewing mode and in the diameter of the central injector tube, i.e. 2.3 for the axially and 1.4 mm for the radially viewed ICP. The figures of merit evaluated were: warm-up time, short- and long-term stabilities, UV and VIS spectral resolution and limit of detection (LOD) for Ni in 0.14 mol l-1 HNO3 and 1000 mg l-1 Cr media, respectively. The influence of residual carbon content on background equivalent concentration (BEC) and sensitivity attained for Al, As, Ca, Cd, Co, Cr, Cu, Fe, Mg, Mn, Se and Zn was also investigated in both viewing modes. The robustness, short- and long-term stabilities, and UV and VIS spectral resolutions were similar for both configurations. For the radially viewed equipment, the warm-up time was at least a factor of 2 shorter than that for the axially viewed configuration. On the other hand, the sensitivity attained for Ni with the axially viewed arrangement was approximately 20 times better than that with the radially viewed. In both viewing modes, and for most of the evaluated elements, the values of BEC and LOD were similar for all studied carbon concentrations (from 10 to 10 000 mg l-1 C) when working with robust plasma conditions. The Mg II/Mg I ratio at an applied power of 1.3 kW and a nebulizer gas flow-rate of 0.90 and 0.70 l min-1 for axially and radially viewed configurations were 10.6 and 13.7, respectively. Quantitative determinations were successfully performed using both systems.

Saved in:
Bibliographic Details
Main Authors: SILVA, F. V., TREVIZAN, L. C., SILVA, C. S., NOGUEIRA, A. R. de A., NÓBREGA, J. A.
Other Authors: FERNANDO V. SILVA, UFSCAR; LILIAN C. TREVIZAN, UFSCAR; CINTIA S. SILVA, UFSCAR; ANA RITA DE ARAUJO NOGUEIRA, CPPSE; JOAQUIM A. NÓBREGA, UFSCAR.
Format: Artigo de periódico biblioteca
Language:Ingles
English
Published: 2003-05-13
Subjects:Figures of merit, Axial viewing, Radial viewing, ICP OES,
Online Access:http://www.alice.cnptia.embrapa.br/alice/handle/doc/46350
https://doi.org/10.1016/S0584-8547(02)00176-3
Tags: Add Tag
No Tags, Be the first to tag this record!
id dig-alice-doc-46350
record_format koha
spelling dig-alice-doc-463502022-06-20T19:19:59Z Evaluation of inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations. SILVA, F. V. TREVIZAN, L. C. SILVA, C. S. NOGUEIRA, A. R. de A. NÓBREGA, J. A. FERNANDO V. SILVA, UFSCAR; LILIAN C. TREVIZAN, UFSCAR; CINTIA S. SILVA, UFSCAR; ANA RITA DE ARAUJO NOGUEIRA, CPPSE; JOAQUIM A. NÓBREGA, UFSCAR. Figures of merit Axial viewing Radial viewing ICP OES The analytical performance of two inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations, equipped with charge coupled device solid-state detectors was evaluated using Ar, Ba, Mg, and Ni as test elements. Both instruments have similar Echelle optical arrangements and radio-frequency generators, differing only in the viewing mode and in the diameter of the central injector tube, i.e. 2.3 for the axially and 1.4 mm for the radially viewed ICP. The figures of merit evaluated were: warm-up time, short- and long-term stabilities, UV and VIS spectral resolution and limit of detection (LOD) for Ni in 0.14 mol l-1 HNO3 and 1000 mg l-1 Cr media, respectively. The influence of residual carbon content on background equivalent concentration (BEC) and sensitivity attained for Al, As, Ca, Cd, Co, Cr, Cu, Fe, Mg, Mn, Se and Zn was also investigated in both viewing modes. The robustness, short- and long-term stabilities, and UV and VIS spectral resolutions were similar for both configurations. For the radially viewed equipment, the warm-up time was at least a factor of 2 shorter than that for the axially viewed configuration. On the other hand, the sensitivity attained for Ni with the axially viewed arrangement was approximately 20 times better than that with the radially viewed. In both viewing modes, and for most of the evaluated elements, the values of BEC and LOD were similar for all studied carbon concentrations (from 10 to 10 000 mg l-1 C) when working with robust plasma conditions. The Mg II/Mg I ratio at an applied power of 1.3 kW and a nebulizer gas flow-rate of 0.90 and 0.70 l min-1 for axially and radially viewed configurations were 10.6 and 13.7, respectively. Quantitative determinations were successfully performed using both systems. 2022-06-20T19:19:51Z 2022-06-20T19:19:51Z 2003-05-13 2002 Artigo de periódico Spectrochimica Acta Part B, v.57, n.12, p.1905-1913, dec. 2002. http://www.alice.cnptia.embrapa.br/alice/handle/doc/46350 https://doi.org/10.1016/S0584-8547(02)00176-3 Ingles en openAccess
institution EMBRAPA
collection DSpace
country Brasil
countrycode BR
component Bibliográfico
access En linea
databasecode dig-alice
tag biblioteca
region America del Sur
libraryname Sistema de bibliotecas de EMBRAPA
language Ingles
English
topic Figures of merit
Axial viewing
Radial viewing
ICP OES
Figures of merit
Axial viewing
Radial viewing
ICP OES
spellingShingle Figures of merit
Axial viewing
Radial viewing
ICP OES
Figures of merit
Axial viewing
Radial viewing
ICP OES
SILVA, F. V.
TREVIZAN, L. C.
SILVA, C. S.
NOGUEIRA, A. R. de A.
NÓBREGA, J. A.
Evaluation of inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations.
description The analytical performance of two inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations, equipped with charge coupled device solid-state detectors was evaluated using Ar, Ba, Mg, and Ni as test elements. Both instruments have similar Echelle optical arrangements and radio-frequency generators, differing only in the viewing mode and in the diameter of the central injector tube, i.e. 2.3 for the axially and 1.4 mm for the radially viewed ICP. The figures of merit evaluated were: warm-up time, short- and long-term stabilities, UV and VIS spectral resolution and limit of detection (LOD) for Ni in 0.14 mol l-1 HNO3 and 1000 mg l-1 Cr media, respectively. The influence of residual carbon content on background equivalent concentration (BEC) and sensitivity attained for Al, As, Ca, Cd, Co, Cr, Cu, Fe, Mg, Mn, Se and Zn was also investigated in both viewing modes. The robustness, short- and long-term stabilities, and UV and VIS spectral resolutions were similar for both configurations. For the radially viewed equipment, the warm-up time was at least a factor of 2 shorter than that for the axially viewed configuration. On the other hand, the sensitivity attained for Ni with the axially viewed arrangement was approximately 20 times better than that with the radially viewed. In both viewing modes, and for most of the evaluated elements, the values of BEC and LOD were similar for all studied carbon concentrations (from 10 to 10 000 mg l-1 C) when working with robust plasma conditions. The Mg II/Mg I ratio at an applied power of 1.3 kW and a nebulizer gas flow-rate of 0.90 and 0.70 l min-1 for axially and radially viewed configurations were 10.6 and 13.7, respectively. Quantitative determinations were successfully performed using both systems.
author2 FERNANDO V. SILVA, UFSCAR; LILIAN C. TREVIZAN, UFSCAR; CINTIA S. SILVA, UFSCAR; ANA RITA DE ARAUJO NOGUEIRA, CPPSE; JOAQUIM A. NÓBREGA, UFSCAR.
author_facet FERNANDO V. SILVA, UFSCAR; LILIAN C. TREVIZAN, UFSCAR; CINTIA S. SILVA, UFSCAR; ANA RITA DE ARAUJO NOGUEIRA, CPPSE; JOAQUIM A. NÓBREGA, UFSCAR.
SILVA, F. V.
TREVIZAN, L. C.
SILVA, C. S.
NOGUEIRA, A. R. de A.
NÓBREGA, J. A.
format Artigo de periódico
topic_facet Figures of merit
Axial viewing
Radial viewing
ICP OES
author SILVA, F. V.
TREVIZAN, L. C.
SILVA, C. S.
NOGUEIRA, A. R. de A.
NÓBREGA, J. A.
author_sort SILVA, F. V.
title Evaluation of inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations.
title_short Evaluation of inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations.
title_full Evaluation of inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations.
title_fullStr Evaluation of inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations.
title_full_unstemmed Evaluation of inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations.
title_sort evaluation of inductively coupled plasma optical emission spectrometers with axially and radially viewed configurations.
publishDate 2003-05-13
url http://www.alice.cnptia.embrapa.br/alice/handle/doc/46350
https://doi.org/10.1016/S0584-8547(02)00176-3
work_keys_str_mv AT silvafv evaluationofinductivelycoupledplasmaopticalemissionspectrometerswithaxiallyandradiallyviewedconfigurations
AT trevizanlc evaluationofinductivelycoupledplasmaopticalemissionspectrometerswithaxiallyandradiallyviewedconfigurations
AT silvacs evaluationofinductivelycoupledplasmaopticalemissionspectrometerswithaxiallyandradiallyviewedconfigurations
AT nogueiraardea evaluationofinductivelycoupledplasmaopticalemissionspectrometerswithaxiallyandradiallyviewedconfigurations
AT nobregaja evaluationofinductivelycoupledplasmaopticalemissionspectrometerswithaxiallyandradiallyviewedconfigurations
_version_ 1756028359336263680