Development and validation of a stability-indicating HPLC method for the determination of buclizine hydrochloride in tablets and oral suspension and its application to dissolution studies

A method using liquid chromatography has been developed and validated for determination of buclizine in pharmaceutical formulations and in release studies. Isocratic chromatography was performed on a C18 column with methanol:water (80:20 v/v, pH 2.6) as mobile phase, at a flow rate of 1.0 mL/min, and UV detection at 230 nm. The method was linear, accurate, precise, sensible and robust. The dissolution test was optimized and validated in terms of dissolution medium, apparatus agitation and rotation speed. The presented analytical and dissolution procedures can be conveniently adopted in the quality and stability control of buclizine in tablets and oral suspension.

Saved in:
Bibliographic Details
Main Authors: Kuminek,Gislaine, Stulzer,Hellen K., Tagliari,Monika P., Oliveira,Paulo R., Bernardi,Larissa S., Rauber,Gabriela, Cardoso,Simone G.
Format: Digital revista
Language:English
Published: Sociedade Brasileira de Química 2012
Online Access:http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000100036